Tektronix TDS644B Caulking Gun User Manual


 
Command Descriptions
2–104
TDS Family Oscilloscope Programmer Manual
In extended-acquisition-length mode, <NR1> ranges up to the extended
acquisition length as opposed to the waveform record length.
DATA:STOP 15000
specifies that the waveform transfer will stop at data point 15000.
DATA:STOP?
might return 14900 as the last data point that will be transferred.
DATa:TARget
Sets or queries the location for storing waveform data transferred to the
instrument using the CURVe command. This command is equivalent to the
DATa:DESTINATION command and is included for compatibility with older
Tektronix instruments.
Waveform
CURVe
DATa:TARget REF<x>
DATa:TARget?
DATa
:
TARget
?
<Space> <x>
REF
DATa:WIDth
Sets the number of bytes per data point in the waveform transferred using the
CURVe command.
Waveform
CURVe, WFMPre:BIT_Nr, WFMPre:BYT_Nr
DATa:WIDth <NR1>
DATa:WIDth?
Examples
Group
Related Commands
Syntax
Group
Related Commands
Syntax