Tektronix TDS644B Caulking Gun User Manual


 
Command Descriptions
2–48
TDS Family Oscilloscope Programmer Manual
When extended-acquisition-length mode is on, this command can still set and
return values. However, this acquire:autosave feature will not actually work
while extended-acquisition-length mode is on.
Acquisition
ACQuire:AUTOSAve { OFF | ON | <NR1> }
ACQuire:AUTOSAve?
ON
ACQuire
:
AUTOSAve
<Space>
?
<NR1>
OFF
OFF or <NR1> = 0 turns autosave off.
ON or <NR1> 0 turns autosave on.
ACQUIRE:AUTOSAVE 1
turns autosave mode on.
ACQUIRE:AUTOSAVE OFF
turns autosave mode off.
ACQUIRE:AUTOSAVE?
might return 1, indicating that autosave mode is on.
ACQuire:MODe
Sets or queries the acquisition mode of the digitizing oscilloscope. This affects
all live waveforms. This command is equivalent to setting Mode in the Acquire
menu.
Waveforms are the displayed data point values taken from acquisition intervals.
Each acquisition interval represents a time duration set by the horizontal scale
(time per division). The digitizing oscilloscope sampling system always samples
at the maximum rate, and so an acquisition interval may include more than one
sample.
The acquisition mode, which you set using this ACQuire:MODe command,
determines how the final value of the acquisition interval is generated from the
many data samples.
Group
Syntax
Arguments
Examples