Tektronix TDS644B Caulking Gun User Manual


 
Command Descriptions
TDS Family Oscilloscope Programmer Manual
2–83
CURSor:HBArs:SELect
Selects which horizontal bar cursor is active for front-panel control. The active
cursor will be displayed as a solid horizontal line and can be moved using the
front-panel general purpose knob when the cursor menu is active. The unselected
cursor will be displayed as a dashed horizontal line. This command is equivalent
to pressing the SELECT button on the front panel when the Cursor menu is
displayed.
Cursor
CURSor:HBArs:SELect { CURSOR1 | CURSOR2 }
CURSor:HBArs:SELect?
CURSor
:
HBArs
:
SELect
<Space>
?
CURSOR2
CURSOR1
CURSOR1 selects the first horizontal bar cursor.
CURSOR2 selects the second horizontal bar cursor.
CURSOR:HBARS:SELECT CURSOR1
selects the first horizontal bar cursor as the active cursor.
CURSOR:HBARS:SELECT?
returns CURSOR1 when the first cursor is the active cursor.
CURSor:HBArs:UNITS
Sets or queries the units for the horizontal bar cursors. This command is
equivalent to setting Amplitude in the Cursor menu.
Cursor
CURSor:HBArs:UNIts { BASe | IRE }
CURSor:HBArs:UNITS?
Group
Syntax
Arguments
Examples
Group
Syntax