Tektronix TDS644B Caulking Gun User Manual


 
Command Descriptions
2–136
TDS Family Oscilloscope Programmer Manual
FACtory (No Query Form)
Resets the digitizing oscilloscope to its factory default settings. This command is
equivalent to selecting Recall Factory Setup in the Waveform Save/Recall
menu.
Miscellaneous
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp
FACtory
FACtory
Setting the digitizing oscilloscope to factory default includes:
H Clears the Event Status Enable Register.
H Clears the Service Request Enable Register.
H Sets the Device Event Status Enable Register to 255.
H Sets the Power On Status Clear Flag to TRUE
H Purges all defined aliases.
H Enables all Command Headers (HEADer ON).
H Sets the macro defined by *DDT to a “zero-length field.”
H Clears the pending operation flag and associated operations.
The FACtory command does not alter the following:
H The state of the GPIB (IEEE Std 488.1-1987) interface.
H The selected GPIB address.
H Calibration data that affects device specifications.
H Protected user data.
H Stored settings.
H The current password (if implemented).
Group
Related Commands
Syntax