Tektronix THS710A Work Light User Manual


 
Specifications
THS710A, THS720A, THS730A & THS720P Service Manual
1–13
Table 1–3: General specifications (Cont.)
Certifications and Compliances
Certifications Listed UL3111-1 and CAN/CSA-C22.2 No. 1010.1-92, complies with
EN61010-1 /A2
O
v
er
voltag
e
C
at
e
go
ry
Category Examples
CAT III A typical CAT III environment is
the power distribution system
within a building or factory. These
environments are somewhat
protected from lightning strikes,
but susceptible to switching
transients and other disturbances
that may generate high voltage
impulses.
CAT II A typical CAT II environment is
the 120/240 V distribution system
within a lab or office. These
environments are fairly well
protected from external high
voltage disturbances.
CAT I A typical CAT II environment is
circuitry inside electrical or elec-
tronic equipment that is powered
by a power supply or a battery.
EC Declaration of
Conformity
Meets the intent of Directive 89/336/EEC for Electromagnetic
Compatibility and Low-Voltage Directive 73/23/ECC for Product Safety
as ammended by 93/68/EEC. Compliance was demonstrated to the
following specifications as listed in the official Journal of the European
Communities:
EN 55011 Class A: Radiated and Conducted Emissions
13
EN 50081–1 Emissions:
EN 60555–2 Power Harmonics
EN 50082–1 Immunity:
IEC 801–2 Electrostatic Discharge
IEC 801–3 RF Radiated
2
IEC 801–4 Fast Transients
IEC 801–5 Surge
3
EN 61010–1 Safety
1
Tektronix-supplied ferrite bead required on instrument end of
RS-232 cable
2
Performance criteria: 5.0 div increase in peak to peak noise
(Sample acquisition mode, full bandwidth); otherwise, 1.0 div
increase in peak to peak noise
3
Applies to instrument operating with Tektronix-supplied AC
adapter