Tektronix THS710A Work Light User Manual


 
Operator Information
THS710A, THS720A, THS730A & THS720P Service Manual
2–9
To maintain signal fidelity, you must compensate each voltage probe for the
channel input it is connected to.
1. Connect the oscilloscope probe and then press AUTORANGE.
Probe tip to PROBE COMP
Probe reference lead to
Ch 1
AUTORANGE
2. Check the shape of the displayed waveform.
Compensated correctly
Overcompensated
Undercompensated
3. If necessary, adjust the probe for correct compensation.
P6117 P5102
4. Repeat these steps for the other probe and channel.
Compensating the
Oscilloscope Probes