Tektronix THS720A Work Light User Manual


 
Maintenance
THS710A, THS720A, THS730A & THS720P Service Manual
6–29
H Displayed information is interrupted by static horizontal or vertical lines
or stripes
3. If (after a short wait) you can see that the display operates normally but the
display backlight does not turn on, then skip to Backlight on page 6–32.
4. If the instrument does not appear to turn on (no display backlight, no sound
of relay clicks, no displayed information), then continue with step 5.
5. Disconnect the AC adapter from the instrument. Remove the front cover
using the procedure Front Cover on page 6–11.
6. Reinstall the battery (removed in the Front Cover procedure).
7. Verify that the battery connector is properly connected. See Figure 6–17.
CAUTION. To avoid damage to the battery or battery contacts, be careful not to
short the conductors at the battery connector.
Test here
Figure 6–17: Battery connector location