Teledyne 7500ZA Welding Consumables User Manual


 
NDIR Gas Analyzer Specifications
Teledyne Analytical Instruments 89
9. Specifications
9.1 General Specifications
9.1.1 Standard Specifications
Measurement Method:
NO, SO
2
, CO
2
, CO, CH
4
:
Non-dispersion infrared-ray absorption
method
Single light source and single beams (single
beam system)
O
2
:
Fuel cell O
2
analyzer (built-in) or
paramagnetic O
2
analyzer (built-in) or
zirconia O
2
analyzer (externally installed
TYPE: ZFK7)
Measured Gas/Range:
Min Range Max Range
NO
0-200 ppm 0-5000ppm
SO
2
0-200 ppm 0-10 vol.%
CO
2
0-100 ppm 0-100 vol.%
CO
0-200 ppm 0-100 vol.%
CH
4
0-500 ppm 0-100 vol.%
O
2
(built in)
Fuel Cell
0-10 vol.% 0-25 vol.%
O
2
(built in)
Paramag.
0-5 vol.% 0-100 vol.%
None 100-95 vol.%