Alarm Setting Descriptions
The alarm contact number can be assigned to any channel.
Channels can also have multiple alarms assigned to them.
Channel: Channel setting targeted for alarm.
H-Limit value: Sets the high limit value (concentration) for
alarm.
L-Limit value: Sets the low limit value (concentration) for alarm.
Kind of Alarm: Select either High limit alarm, Low limit alarm,
and High limit or Low limit alarm, HH limit
alarm, and LL limit alarm.
High, HH: Alarm contact closes when above
H-limit alarm.
Low, LL: Alarm contact closes when below
L-limit alarm.
High or Low: Alarm contact closes when above
H-limit value or below lower limit
value.
ON/OFF: Enables the alarm function when set to ON, or
disables it if set to OFF.
Note: The H-limit value cannot be set below the L-limit value, and
the L-limit value cannot be set above the H-limit value.
If it is desired to set the H-limit value below the L-limit
value already stored in the memory, reduce the L-limit
value beforehand, and vice versa for setting the low limiot
above the high limit.
Note: After turning on power, the alarm logic trigger is inactive for
10 minutes.
When an H-limit alarm occurs, the
“H-alarm” message displays in the relevant
component field. Other alarm displays are:
L-alarm for Low-limit alarm, HH-
alarm for high limit alarm, HH-alarm for
high-high alarm, and LL-alarm for low-
low alarm.