Teledyne 7500ZA Welding Consumables User Manual


 
Installation Model 7500ZA
Teledyne Analytical Instruments 10
Entry of dust into the instrument may result in poor performance.
Use clean tubing and couplings.
Sample inlet: Attach clean, dry sample gas to the inlet port.
Note that the calibration gases ( zero and span)
also connect to this port.
Input gas flow t should be constant within the
range of 0.5 L/min ±0.2 L/min.
Sample outlet: Attach tubing for exhausting sample or calibration
gases. The exhaust line should be directed to an
appropriate safe vent depending on the nature of
the sample/calibration gas used.
Purge gas inlet: Used to connect a purge gas for purging the
internal housing of the analyzer. Use dry N
2
or
instrument air for purge gas. (Flow rate is 1L/min
or more. Keep dust or moisture/mist from
entering the purge line.
CAUTION: THIS INSTRUMENT IS DESIGNED TO HANDLE
HAZARDOUS GASES. WHENEVER A FITTING IS
OPENED, USE A NEW FERRULE AND CONE TO
SECURE A GAS TIGHT SEAL. EACH FITTING MUST
BE LEAK CHECKED WHENEVER A CONNECTION
HAS BEEN OPENED OR DISTURBED IN ANY
MANNER.
The internal piping diagram for the Model 7500ZA is shown
below.