HP (Hewlett-Packard) 16501A Sander User Manual


 
HTIMe
Query :INTermodule:HTIMe?
The HTIMe query returns a value representing the internal hardware skew in
the Intermodule configuration. If there is no internal skew, 9.9E37 is
returned.
The internal hardware skew is only a display adjustment for time-correlated
waveforms. The value returned is the average propagation delay of the trigger
lines through the intermodule bus circuitry. These values are for reference only
because the values returned by TTIMe include the internal hardware skew
represented by HTIMe.
Returned Format
[:INTermodule:HTIMe] <value_A>,<value_B>,<value_C>,<value_D>,
<value_E>[,<value_F>,<value_G>,<value_H>,<value_I>,<value_J>]
<NL>
<value_X> Skew for module in slot X (real number)
Example OUTPUT XXX;":INTERMODULE:HTIME?"
INTermodule Subsystem
HTIMe
13–7