CALIBRATION PROCEDURE
Calibration parameters are logically divided into groups
of two parameters each – initial scale value and final
scale value. A calibration check is provided after entering
the values of each group. It is also possible to perform a
calibration check without making an entry: press the
FUNC button twice when “OFF” is displayed. The instru-
ment goes directly to the group check.
Before beginning calibration, be sure the internal DIP
switch V101 is open.
WARNING: Perform the calibration procedure according
to J106 jumper positions as shown in Figure 1. Other-
wise, the stored calibration values may be lost.
epyTtupnI
601J
2-14-36-58-7
,DTR,C/T
Vm06-21,Vm06-0
)noitarugifnocdradnats(
esolc
nepo
nepo nepo
V5-1,V5-0
nepo esolc nepo nepo
V01-2,V01-0 nepo nepo esolc nepo
Am02-4,Am02-0
nepo nepo nepo esolc
K081=291RCTTCtupnI
nepO=101HCdesolC=021HS
Figure 1. J106 Settings
General Guidelines
a) The instrument should be mounted in its case in order
to keep the internal temperature constant.
b) Ambient temperature should be stable. Avoid drift due
to air conditioning or other mechanical devices.
c) Relative humidity should not exceed 70%.
d) Minimum warm up time of at least 20 minutes.
e) Operate as much as possible in a noise free
environment.
f) during calibration, connect one input at a time to the
rear terminal block.
g) Use calibrators with the following:
Accuracy
TC Input: +0.005% output
+0.001% range
+5 mV
RTD Input: +0.02%
+0.0025 decade
CJ Compensation: better than 0.1°C
Current Transformer: 0.1 mA AC rms
Resolution
TC Input: 1 mV
RTD Input: 10 m
CJ Compensation: better than 0.1°C
Current Transformer: 0.1 mA AC rms
Calibration Parameters
Following is a complete list of calibration symbols:
CodeParameter
tL TC Input Initial Scale Value
tH TC Input Final Scale Value
t. TC Input Check
rJ Cold Junction Compensation
rJ. Cold Junction Compensation Check
PL RTD Input Initial Scale Value
PH RTD Input Final Scale Value
P. RTD Input Check
AL Current Input Initial Scale Value
AH Current Input Final Scale Value
A. Current Input Check
nL 5 Volt Input Initial Scale Value
nH 5 Volt Input Final Scale Value
n. 5 Volt Input Check
UL 10 Volt Input Initial Scale Value
UH 10 Volt Input Final Scale Value
U. 10 Volt Input Check