Tektronix TDS 500D Sander User Manual


 
Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–91
Optical Filters Checks (TDS 500D/TDS 700D Only)
The procedure that follows verifies that the frequency response of the combined
P670xB and TDS500D/700D oscilloscope system is a Fourth Order Bessel
Thompson Filter which is within limits.
Nominal Filters are available with Option 2C. Reference Receiver Filters are
available with Options 3C and 4C. Options 3C and 4C are not available without
2C, and 3C and 4C are not available on the TDS 794D.
You can verify Option 2C is enabled on the TDS 794D by pressing MEASURE,
then selecting Masks from the Measure pop-up menu. Table 1–8 lists the
available vertical scale factors for each option. Tables 1–9, 1–10, and 1–11 list
the available filters and their specifications.
Table 1–8: Available Filters
Nominal Filters
(Option 2C)
Reference Receivers Filters
(Options 3C, 4C
)
1 mW per division
2 mW per division
5 mW per division
10 mW per division 10 mW per division
20 mW per division 20 mW per division
50 mW per division 50 mW per division
100 mW per division
200 mW per division
500 mW per division
1 mW per division
Table 1–9: Reference Receiver Filter Options:
Option 4C – SONET (P6703B 1300nm) Option 3C – Fibre Channel (P6701B 850nm)
52Mbit OC1 FC133Mbit
FC266Mbit
FC531Mbit
155Mbit OC3 155Mbit OC3
622Mbit OC12 622Mbit OC12
FC1063 (TDS 784D Only) FC1063 (TDS 784D Only)
Filter Availability