Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–75
b. Press WAVEFORM OFF.
c. Press HORIZONTAL MENU.
d. Use the keypad to set horizontal scale (/div) to 50 ms (press 50, SHIFT,
m, then ENTER).
e. Press SHIFT; then press ACQUIRE MENU.
f. Press the main-menu Stop After.
g. Press the side-menu RUN/STOP button only.
h. Press the main-menu Mode.
i. Press the side-menu Sample.
j. Press RUN/STOP.
k. Press VERTICAL MENU.
l. Use the keypad to set fine scale to 300 mV (press 300, SHIFT, m, then
ENTER).
m. CONFIRM that the TRIG’D LED stays lighted and that the waveform
on screen is stable. Also, CONFIRM that the waveform on the screen
has one positive pulse and a number of negative pulses. See Figure 1–29.
Positive pulse
Negative pulses
Figure 1–29: Triggered signal range test – 300 mV