Tektronix TDS 500D Sander User Manual


 
Performance Tests
1–56
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
d. Check against limits: Do the following subparts in the order listed.
H CHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
Table 1–7: Delayed trigger level accuracy readout
Instrument Minimum Maximum
TDS 600B 9.863 V 10.137 V
TDS 500D/700D 9.8853 V 10.1147 V
TDS 794D 4.8953 V 5.1047 V
H Enter voltage on test record.
H Press the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Press the main-menu button
Level. Repeat substep c.
H CHECK that the Level readout in the side menu is within the limits
listed in Table 1–7.
H Enter voltage on test record.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Disconnect the cable from the generator output at the input connector of
CH 1.