Agilent Technologies 6813B Welding System User Manual


 
Language Dictionary - 3
29
Calibration Subsystem Commands
The commands in this subsystem allow you to do the following:
u Enable and disable the calibration mode
u Change the calibration password
u Calibrate the current and voltage output levels, and store new calibration constants in nonvolatile
memory.
Subsystem Syntax
CALibrate
:CURRent
:AC Begin ac current programming calibration sequence
:MEASure Begin current measurement calibration sequence
:DATA <n> Input a calibration measurement
:IMPedance Begin output impedance calibration sequence
:LEVel <level> Advance to next calibration step (P1 | P2 | P3 | P4)
:PASSword <n> Set calibration password
:PWM
:FREQuency <n> Trim pulse width modulator frequency
:RAMP <n> Trim pulse width modulator ramp
:SAVE Save new cal constants in non-volatile memory
:STATE <bool> [,<n>] Enable or disable calibration mode
:VOLTage
:AC Begin ac voltage calibration sequence
:DC Begin dc voltage calibration sequence
:OFFSet Begin offset voltage calibration sequence
:PROTection Begin voltage protection calibration sequence
:RTIMe Begin realtime voltage calibration sequence
CALibrate:CURRent:AC
Phase Selectable
This command can only be used in the calibration mode. It initiates the calibration of the ac current limit
and metering circuits.
Command Syntax
CALibrate:CURRent:AC
Parameters
None
Examples
CAL:CURR:AC
Related Commands
CAL:STAT CAL:SAV CAL:LEV