Agilent Technologies 6813B Welding System User Manual


 
3 - Language Dictionary
32
CALibrate:STATe
This command enables and disables calibration mode. The calibration mode must be enabled before the
will accept any other calibration commands. The first parameter specifies the enabled or disabled state.
The second parameter is the password. It is required if the calibration mode is being enabled and the
existing password is not 0. If the password is not entered or is incorrect, an error is generated and the
calibration mode remains disabled. The query statement returns only the state, not the password.
Whenever the calibration state is changed from enabled to disabled, any new calibration constants are
lost unless they have been stored with CALibrate:SAVE.
Command Syntax
CALibrate:STATe <bool> [,<NRf>]
Parameters
0 | 1 | OFF | ON [,<password>]
*RST Value
OFF
Examples
CAL:STAT 1,6812 CAL:STAT OFF
Query Syntax
CALibrate:STATe?
Returned Parameters
<NR1>
Related Commands
CAL:PASS CAL:SAVE
CALibrate:VOLTage:AC
Phase Selectable
This command can only be used in calibration mode. It initiates the calibration of the ac voltage
programming and metering circuits.
Command Syntax
CALibrate:VOLTage:AC
Parameters
None
Examples
CAL:VOLT:AC
Related Commands
CAL:SAVE CAL:STAT
CALibrate:VOLTage:DC
Agilent 6811B, 6812B, 6813B, Only
This command can only be used in calibration mode. It initiates the calibration of the dc voltage
programming circuits.
Command Syntax
CALibrate:VOLTage:DC
Parameters
None
Examples
CAL:VOLT:DC
Related Commands
CAL:SAVE CAL:STAT