3.11 Screen/Display Functions
3.11.1 Brightness control
The DISPlay:BRIGhtness command allows you to control the brightness of the
trace(s) displayed on the screen of your CombiScope instrument on a scale from
0.0 (low) to 1.0 (high). After a
*
RST command, the brightness intensity is 0.18.
PROGRAM EXAMPLE:
CALL Send(0, 8, "DISPlay:BRIGhtness .3", 1)
Sets brightness at 0.3
.
3.11.2 Display functions
The DISPlay:WINDow and DISPlay:MENU commands allow you to use the
following display functions:
The WINDow1 functions use the front panel screen display of MEAS1/MEAS2,
CURSORS, and MATH-FFT to read measurement data from the CombiScope
instrument (refer to section 3.11.2.1).
The WINDow2 function to write user-defined text on the screen (refer to
section 3.11.2.2).
The MENU function to display softkey menus on the screen (refer to
section 3.11.2.3).
The layout of the display areas on the screen is as follows:
WINDow[1]
MENU