Sterling 882.01749.00 Welder User Manual


 
882.01749.00 37 of 93
Screw Recovery Time.
The screw recovery time is the estimated screw recovery time, and is used as the initial value for
validating the recipe. Actual screw recovery times as measured by the presence of the screw recovery signal,
are used to fine-tune the feed rate during operation. The estimated time must be at least ½ of the actual
recovery time. If less than ½, the feeder will interpret the recovery signal as a purge cycle.
Calibration weight.
Calibration weight is determined from chart 5.1, and is dependant on the disc installed in the feeder.
This weight is the amount of standard density material dispensed by the feeder during one revolution of the
disc. This weight is used as the initial value for recipe validation and feed rate calculation. If samples are not
weighed as described in the next section, this value is used during operation. The weight can be entered in
ounces or grams. The unit of measure is changed by pressing the units button at the middle right of the screen.
Changing the unit of measure will NOT change the value displayed on the screen. The value initially displayed
is the weight which would result in a motor speed of 25%. The valid range of values would then be ¼ to 5
times this initial value.
Additive Sample weighing.
Weighing samples of additive provides for more precise control of feed rate. Up to 5 samples can be
weighed. All weights entered are averaged to arrive at a value to be used during operation. The weighing
process may be skipped altogether, in which case the calibration weight is used for calculations of feed rate.
See the section Weight Calibration at the end of this section.