Tektronix TDS 700D Sander User Manual


 
Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–39
Table 1–4: Analog bandwidth
er ical cale
Re ere ce ampli e
ri al
Test frequency
–3 b Limi
V
er
t
ical
s
cale
Re
f
ere
n
ce
ampli
tud
e
(6 divisions)
Ho
ri
zont
al
scale
TDS 794D
TDS 580D
TDS 680B
TDS 684B
TDS 784D
TDS 620B
TDS 644B
TDS 520D
TDS 540D
TDS 724D
TDS 754D
–3
d
b Limi
ts
100 mV 600 mV 1 ns 2 GHz 1 GHz 500 MHz 500 MHz 424 mV
1mV 6mV 1 ns NA 500 MHz 250 MHz 450 MHz 4.24 mV
2mV 12 mV 1 ns NA 600 MHz 300 MHz 500 MHz 8.48 mV
5mV 30 mV 1 ns NA 750 MHz 450 MHz 500 MHz 21.2 mV
10 mV 60 mV 1 ns 2 GHz 1 GHz 500 MHz 500 MHz 42.4 mV
20 mV 120 mV 1 ns 2 GHz 1 GHz 500 MHz 500 MHz 84.8 mV
50 mV 300 mV 1 ns 2 GHz 1 GHz 500 MHz 500 MHz 212 mV
200 mV 1.2 V 1 ns 2 GHz 1 GHz 500 MHz 500 MHz 848 mV
500 mV 3 V
1
1 ns 2 GHz 1 GHz 500 MHz 500 MHz 2.12 V
1
1 V 6 V
1
1 ns 2 GHz 1 GHz 500 MHz 500 MHz 4.24 V
1
1
If your generator cannot output 6 divisions of amplitude, set it to its maximum output. The –3 db limit can be calculated as:
0.707 reference amplitude.
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
H Press MEASURE; then press the main-menu button Select
Measrmnt for CHx.
H Press the side-menu button more, if needed, until the menu label
Frequency appears in the side menu (its icon is shown at the left).
Press the side-menu button Frequency.
H Press the side-menu button more until the menu label Pk-Pk
appears in the side menu (its icon is shown at the left). Press the
side-menu button Pk-Pk.
H Press CLEAR MENU.
H Set the generator output so the CHx Pk-Pk readout equals the
reference amplitude in Table 1–4 that corresponds to the vertical
scale set in substep d.
H Press the front-panel button SET LEVEL TO 50% as necessary to
trigger a stable display. At full bandwidth, you may also want to
make small, manual adjustments to the trigger level. You can use the
TRIGGER LEVEL knob to do this. (Full bandwidth varies with
TDS model as is shown in Table 1–4).