Tektronix TDS 700D Sander User Manual


 
Performance Tests
1–44
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
H Press CURSOR.
H Press the main-menu button Function; then press the side-menu
button V Bars.
H Press CLEAR MENU.
H Align one V bar cursor to the time reference point of the left-most
waveform edge and the other cursor to the time reference point of the
right-most waveform edge by rotating the General Purpose knob.
(Press SELECT to switch between the two cursors). See Figure
1–11 on page 1–44.
H Read the measurement results at the D: cursor readout, not the @:
readout on screen.
Locate the time reference
points for these waveforms.
2
Display the waveforms.
1
Read results.
4
Align each cursor to the time
reference points
3
Figure 1–11: Measurement of channel delay – TDS 684B shown
g. Check against limits: CHECK that the cursor readout on screen is
100 ps for the TDS 600B or 50 ps for the TDS 500D/700D.
h. If the channel skew is within the limits, enter time on the test record and
proceed to step 3. Otherwise, proceed with steps i through p.