Teledyne 7600 Welding Consumables User Manual


 
Infrared Gas Analyzer Operation
Teledyne Analytical Instruments 59
5. In the “Setting of Auto Calibration” screen that appears,
point the cursor to any item you want to set by pressing
the
UP or DOWN key. Press the ENT key.
6. In the “Auto Calibration Parameter Setting” screen that
appears, perform the value entry or the setting. For the
value entry or setting change, use the UP or DOWN key.
To change the setting, use the SIDE key to move the
cursor to the right.
7.
After changing the setting, press the ENT key to accept
the change. Auto calibration will now be carried out
using the entered setting value.
To close the “Setting of Auto Calibration” or cancel the mode
midway, press the ESC key. A previous screen will return.
3.5.4.2 GAS FLOW TIME SETTING
In this submenu you select the gas whose flow time is to be
changed. The flow time range is from 60 to 900 seconds. The initial
value is set for 300 seconds.