Teledyne 7600 Welding Consumables User Manual


 
Operation Model 7600
Teledyne Analytical Instruments 62
When a channel enabled for auto calibration is in the calibration
mode, a message indicating ZERO cal. or SPAN cal. will appear on the
measurement screen. For example, if channels 1 and 2 have the auto
calibration setting enabled, when the calibration event takes place, both
channels 1 and 2 will indicate ZERO cal. when the zero calibration is
being performed. Zero calibrations are performed first and all enabled
channels will zero simultaneously.
When the zero calibrations are finished, channel 1 will enter the
span calibration followed by channel 2.
3.5.4.4 REMOTE START
An auto calibration is always available by keeping the remote start
input closed for at least 1.5 seconds. The remote start can initiate an auto
calibration whether the parameter setting is ON or OFF.
3.5.4.5 FORCED RUN/STOP OF AUTO CALIBRATION
Auto calibration can be performed manually or forcibly stopped
while the calibration is performed.