Teledyne 7600 Welding Consumables User Manual


 
Operation Model 7600
Teledyne Analytical Instruments 74
Reset Av. Output: Resets the average value.
Response time: Sets the response time of the electrical system.
Average Period: Sets the moving average time.
Backlight Timer: Sets an automatic OFF of the backlight of display
unit and the time until the backlight is turned OFF.
Maintenance mode: Enters passwords to switch to the Maintenance
mode.
For the maintenance mode, see Section 3.6.
To make parameter changes:
1.
From the measurement mode, enter the user mode by
pressing the MODE key.
2.
Point the cursor to “Parameter Setting” by pressing the
UP and DOWN
keys. Then press ENT.
3.
In the “Parameter Setting” screen that appears, point the
cursor to any item you want to set by pressing the UP or
DOWN key. Then Press ENT.
4.
Enter numeric values by using the UP or DOWN keys.
Use the SIDE key to move the cursor to the right. After
setting, press the ENT key to save the parameter setting.