Teledyne 7600 Welding Consumables User Manual


 
Operation Model 7600
Teledyne Analytical Instruments 68
3.5.5.2 FORCED RUN/STOP OF AUTO ZERO CALIBRATION
Like the auto calibration, the auto zero calibration can be
performed manually or forcibly stopped while the calibration is
performed.
3.5.5.2.1 MANUAL EXECUTION OF AUTO ZERO CALIBRATION
To perform the auto calibration cycle manually:
1.
From the measurement mode, press the MODE key to
display the user mode.
2.
Point the cursor to “Setting of Auto Zero Calibration” by
pressing the UP or DOWN key and press the ENT key.
3.
In the “Setting of Auto Zero Calibration” item selection
screen that appears, point the cursor to “Auto zero
Calibration Run” by pressing the
UP or DOWN keys.
Press the ENT key.
4.
“Run” is highlighted, displaying a message to confirm the
execution of auto calibration. Press the ENT key to
execute the auto calibration, or press the ESC key to
cancel.