Emerson Process Management BINOS 100 F OXYNOS 100 Welding Consumables User Manual


 
8 - 3
SETTING SYSTEM PARAMETERS
ETC00781(4) Series 100 e 02/2004
8.3 Cross Compensation Calibration (internal)
Determination of cross compensation correction factors is performed during the span adjustment.
Pure test gases are required for this operation. Once cross compensation corrections have been
determined, span adjustments may be performed using test gas mixtures.
Input value 0: spanning without cross compensation correction (test gas mixtures)
Input value 1: spanning with cross compensation correction (pure test gases)
Change values using
and
To perform a calibration with cross compensation correction, proceed as follows:
First perform a zeroing for both analysis channels (see chapter 9.1.1).
Then perform a spanning for both analysis channels as described in chapter 9.1.2.
Repeat spanning for the analysis channel that has been spanned first.
The values entered as described in chapters 8.2 and 8.3 must be “1” for
performing a calibration with cross compensation correction !
Use pure test gases only!
When using test gas mixtures, “C.Cal” must be set to “0” !
CROSS COMPENSATION CALIBRATION (INTERNAL)