SAFETY SUMMARY
S - 4
ETC00781(4) Series 100 e 02/2004
GASES AND GAS CONDITIONING (SAMPLE HANDLING) / SUPPLY VOLTAGE
IV.Gases and Gas Conditioning (Sample Handling)
Be sure to observe the safety regulations for the gases
(sample gas and test gases / span gases) and the gas bottles !
Inflammable or explosive gas mixtures must not be purged into the instrument
without supplementary protective measures !
To avoid risks for the operators by explosive, toxic or unhealthy gas
components, first purge the gas lines with ambient air or nitrogen (N
2
) before
cleaning or exchanging parts of the gas paths.
V. Supply Voltage
The socket outlet shall be installed near the equipment and shall be easily
accessible to disconnect the device from the socket outlet.
Verify whether the line voltage stated on the instrument or power supply is in
accordance with that of your mains line!
Be sure to observe the safety precautions and warnings given by
manufacturer of power supply !
X BINOS
®
100(M), BINOS
®
100 2M (external PS), HYDROS
®
100 and OXYNOS
®
100 are
Safety Class III instruments.
Verify correct polarity for 24 V DC operation !
Use only power supply VSE 2000, UPS 01 T, DP 157, SL5, SL10 (DP 157 and
SL for rack installation only) or equivalent power supplies to keep the
instrument safe.
If using equivalent power supplies they must have SELV output voltage !