Error Code Possible Reasons Check / Correct
16 - 7
LIST OF FAILURES
ETC00781(4) Series 100 e 02/2004
1. Needle valve is not opened
(Option)
2. Sample gas pump (Option)
is not switched on
3. Membrane of pump is
defective
4. Sample gas pump is
defective
5. Solenoid valves defective
(Option)
6. Contamination of the
gas lines
1. Open needle valve
(Option, see Fig. 1-2)
2. Press key PUMP
3. Exchange membrane of the pump.
4. Exchange sample gas pump
5. Check the valve face of the solenoid
valves and exchange if necessary.
Exchange solenoid valves
6. Check gas lines including filters to
contamination.
Clean the gas lines and exchange
the filter elements.
No gas flow