6 - 1
SWITCHING ON
ETC00781(4) Series 100 e 02/2004
6. Switching On
6.1 General
Be sure to observe the safety precautions and warnings !
Once the instrument has been correctly assembled and installed in accordance with the general
instructions given in section 5., the equipment is ready for operation.
The equipment is switched on by providing the required voltage.
The presence of the supply voltage will be indicated by the illumination of the LED displays.
Upon connection of the supply voltage, the analyzer will perform a self diagnostic test routine.
First the actual program version will be shown.
(analyzers without fieldbus) or (analyzers with fieldbus)
Finally either concentration values or error messages will be displayed
If as a result of a battery fault the default values were charged, this will be shown by a flushing “batt.”
This message will disappear after depressing any key.
Analyzer warming-up takes about 15 to 50 minutes, depending on the
installed detectors !
Before starting an analysis, however, the following should be performed:
R entry of the desired system parameters,
R calibration of the analyzer.
Note:
The "X’s" shown in the display indicate a number or combinations of numbers.
GENERAL