Status and Events
48
PS2520G and PS2521G Programmer Manual
Table 8: QUEStionable INSTrument ISUMmary<n> Status
Register
Bit Number
Name Condition (Bit is True)
0 Output<n>
Voltage
Output<n> Constant Current (CC), voltage
unregulated.
1 Output<n>
Current
Output<n> Constant Voltage (CV), current
unregulated.
2 — Not used.
3 — Not used.
4 — Not used.
5 — Not used.
6 — Not used.
7 — Not used.
8 — Not used.
9 Output<n>
OVP
Output<n> overvoltage protection tripped.
10 Output<n>
OCP
Output<n> overcurrent protection tripped.
11 — Not used.
12 — Not used.
13 — Not used.
14 — Not used.
15 — Not used.
The command STATus:QUEStionable:INSTrument:ISUMmaĆ
ry<n>:CONDition? reads the QUEStionable INSTrument ISUMma-
ry<n> CONDition register but does not clear it.
The command STATus:QUEStionable:INSTrument:ISUMmaĆ
ry<n>[:EVENt]? reads and clears the QUEStionable INSTrument
ISUMmary<n> EVENt Status Register.