Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–87
H Disconnect the signal source from CH1, wait a few seconds, then
reconnect the signal.
H CONFIRM that the TRIG’D LED is lighted and the waveform is
stable.
H Disconnect all test equipment from the oscilloscope.
H Press save/recall SETUP, the main-menu button Recall Factory
Setup, and the side-menu OK Confirm Factory Init.