Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–21
TDS 600B Test Record (cont.)
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 600B Performance Test MaximumOutgoingIncomingMinimum
CH 3 or AX1 SIGNAL OUTPUT, 1 MW Pk-Pk ≥ 80 mV __________ __________ Pk-Pk ≤ 120 mV
CH 3 or AX1 SIGNAL OUTPUT, 50 W Pk-Pk ≥ 40 mV __________ __________ Pk-Pk ≤ 60 mV
Probe Compensator Output Signal
Frequency (CH1 Freq). 950 Hz __________ __________ 1.050 kHz
Voltage (difference) 495 mV __________ __________ 505 mV