Tektronix TDS 600B Sander User Manual


 
Performance Tests
1–40
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
f. Measure the test signal:
H Set the frequency of the generator, as shown on screen, to the test
frequency in Table 1–4 that corresponds to the vertical scale set in
substep d. See Figure 1–9.
H Set the horizontal SCALE to the horizontal scale setting in
Table 1–4 that corresponds to the vertical scale set in substep d.
Press SET LEVEL TO 50% as necessary to trigger the signal.
H Read the results at the CHx Pk-Pk readout, which will automatically
measure the amplitude of the test signal. See Figure 1–9.
Set the generator
(reference) frequency to
the test frequency from
Table 1–4.
1
Set the horizontal scale
from Table 1–4.
2
Read results.
3
Figure 1–9: Measurement of analog bandwidth
g. Check against limits:
H CHECK that the Pk-Pk readout on screen is within the limits listed
in Table 1–4 for the current vertical scale setting.
H Enter voltage on test record.
H When finished checking, set the horizontal SCALE back to the
50 ns setting.