Tektronix TDS 600B Sander User Manual


 
Performance Tests
1–80
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
PAL signal
source
Oscilloscope
75 W cable
75 W terminator
Output
Figure 1–35: Line count accuracy test hookup
H Press the main-menu Trigger Position.
H Press the side-menu to Set to 50%.
H Press the main-menu to Horiz Pos.
H Press the side-menu to Set to 50%.
H Use the HORIZONTAL POSITION knob to move the falling edge
of the sync pulse to two divisions to the left of center screen. See
Figure 1–36.