Performance Tests
1–66
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Equipment
required
One female BNC to clip adapter (Item 3)
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
Two precision 50 W coaxial cables (Item 5)
One DC calibration generator (Item 9)
Prerequisites See page 1–15. Also, the oscilloscope must have passed Check
Accuracy For Long-Term Sample Rate, Delay Time, and Delta Time
Measurements on page 1–46.
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal:
H Connect one of the 50 W cables to CH 1. See Figure 1–22.
H Connect the other end of the cable just installed to the female
BNC-to-clips adapter. See Figure 1–22.
H Connect the red clip on the adapter just installed to the PROBE
COMPENSATION SIGNAL on the front panel; connect the black
clip to PROBE COMPENSATION GND. See Figure 1–22.
Oscilloscope
50 W coaxial cable
Female BNC to
clip adapter
Black lead
to GND
Figure 1–22: Initial test hookup
b. Initialize the oscilloscope:
H Press save/recall SETUP.
H Press the main-menu button Recall Factory Setup.
Check Probe
Compensator Output