Tektronix TDS 600B Sander User Manual


 
Performance Tests
1–26
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
TDS 794D Test Record
Photocopy this and the next two pages and use them to record the performance
test results for your TDS 794D.
TDS 794D Test Record
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 794D Performance Test Minimum Incoming Outgoing Maximum
Offset Accuracy
CH1 Offset +10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH2 Offset +10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH3 Offset +10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
CH4 Offset +10 mV
+101 mV
– 2.5 mV
– 25.1 mV
__________
__________
__________
__________
+ 2.5 mV
+ 25.1 mV
DC Voltage Measurement Accuracy (Averaged)
CH1 10 mV Vert scale setting,
–5 Div position setting, +.5 V offset
+0.5755 V __________ __________ +0.5845 V
CH1 10 mV Vert scale setting,
+5 Div position setting, –.5 V offset
–0.5845 V __________ __________ –0.5755 V
CH1 200 mV Vert scale setting,
0 Div position setting, +5 V offset
+5.5345 V __________ __________ +5.6655 V
CH1 200 mV Vert scale setting,
0 Div position setting, –5 V offset
–5.6655 V __________ __________ –5.5345 V
CH1 1 V Vert scale setting,
0 Div position setting, +2.5 V offset
+5.289 V __________ __________ +5.711 V
CH1 1 V Vert scale setting,
0 Div position setting, –2.5 V offset
–5.711 V __________ __________ –5.289 V
CH2 10 mV Vert scale setting,
–5 Div position setting, +.5 V offset
+0.5755 V __________ __________ +0.5845 V
CH2 10 mV Vert scale setting,
+5 Div position setting, –.5 V offset
–0.5845 V __________ __________ –0.5755 V
CH2 200 mV Vert scale setting,
0 Div position setting, +5 V offset
+5.5345 V __________ __________ +5.6655 V