Performance Tests
TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
1–25
TDS 500D/700D Test Record (cont.)
Instrument Serial Number: Certificate Number:
Temperature: RH %:
Date of Calibration: Technician:
TDS 500D/700D Performance Test MaximumOutgoingIncomingMinimum
CH 3 or AX 1 SIGNAL OUTPUT, 1 MW Pk-Pk ≥ 88 mV __________ __________ Pk-Pk ≤ 132 mV
CH 3 or AX 1 SIGNAL OUTPUT, 50 W Pk-Pk ≥ 44 mV __________ __________ Pk-Pk ≤ 66 mV
Probe Compensator Output Signal
Frequency (CH1 Freq). 950 Hz __________ __________ 1.050 kHz
Voltage (difference) 495 mV __________ __________ 505 mV