TDS 500D, TDS 600B & TDS 700D Performance Verification and Specifications
Equipment Required
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduction.
Table 1–1: Test equipment
Item number and
description
Minimum requirements Example Purpose
1. Attenuator,10X
(two required)
Ratio: 10X; impedance 50 W; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0059-02
Signal Attenuation
2. Attenuator, 5X Ratio: 5X; impedance 50 W; connec-
tors: female BNC input, male BNC
output
Tektronix part number
011-0060-02
Signal Attenuation
3. Adapter, BNC female-to-
Clip Leads
BNC female-to-Clip Leads Tektronix part number
013-0076-00
Signal Coupling for Probe
Compensator Output Check
4. Terminator, 50 W Impedance 50 W; connectors: female
BNC input, male BNC output
Tektronix part number
011-0049-01
Signal Termination for
Channel Delay Test
5. Cable, Precision 50 W
Coaxial (two required)
50 W , 36 in, male-to-male BNC
connectors
Tektronix part number
012-0482-00
Signal Interconnection
6. Connector, Dual-Banana
(two required)
Female BNC-to-dual banana Tektronix part number
103-0090-00
Various Accuracy Tests
7. Connector, BNC “T” Male BNC-to-dual female BNC Tektronix part number
103-0030-00
Checking Trigger Sensitivity
8. Coupler, Dual-Input Female BNC-to-dual male BNC Tektronix part number
067-0525-02
Checking Delay Between
Channels
9. Generator, DC Calibra-
tion
Variable amplitude to ±104 V; accura-
cy to 0.1%
Data Precision 8200
1
Checking DC Offset, Gain,
and Measurement Accuracy
10. Generator, Calibration 500 mV square wave calibrator
amplitude; accuracy to 0.25%
Tegam/Tektronix PG 506A
Calibration Generator
1,2
To check accuracy of CH 3
Signal Out
11. Generator, Time Mark
(optional)
Variable marker frequency from 10 ms
to 10 ns; accuracy within 2 ppm
Tegam/Tektronix TG 501A
Time Mark Generator
1,2
Checking Sample-Rate and
Delay-time Accuracy
12. Probe, 10X A P6139A, P6243, P6245, or P6339A
probe
3
Tektronix part number
P6139A or P6245
Signal Interconnection
13. 3.5 inch, 720 K or
1.44 Mbyte, DOS-com-
patible floppy disk
TDS Family Programmer Disk
(Tektronix part number
063–3002–00), included with
User Manual (Tektronix part
number 071–0130-XX)
Checking File System Basic
Functionality
14. Generator, Sine Wave 250 kHz to at least 500 MHz (higher
for higher-bandwidth scopes). Vari-
able amplitude from 60 mV to 2 V
p-p
into 50 W. Frequency accuracy >2.0%
Rohde & Schwarz SMT or
SMY
1,4
Checking Analog Bandwidth,
Trigger Sensitivity, Sample-
rate, External Clock, and
Delay-Time Accuracy