Tektronix THS720P Work Light User Manual


 
THS710A, THS720A, THS730A & THS720P Service Manual
1–1
Specifications
This chapter contains the oscilloscope, DMM, and general specifications for the
THS710A, THS720A, THS730A, and THS720P TekScope instruments. All
specifications are guaranteed unless noted “typical.” Typical specifications are
provided for your convenience but are not guaranteed. Specifications that are
marked with the n symbol are checked in the chapter Performance Verification.
All specifications apply to all four instruments unless noted otherwise. All
specifications assume horizontal MAG is off, unless noted otherwise. To meet
specifications, two conditions must first be met:
H The TekScope instrument must have been operating continuously for ten
minutes within the operating temperature range specified.
H You must perform the Compensate Signal Path operation, accessible through
the utility menu, if the ambient temperature changes by more than 5° C.
Table 1–1: Oscilloscope specifications
Acquisition
Acquisition Modes Sample (Normal), Peak detect, Envelope, and Average
Acquisition Rate,
typical
Up to 25 waveforms per second (2 channels, sample acquisition mode,
MAG on, no measurements)
S
ingl
e
Seq
u
e
nc
e
Acquisition Mode Acquisition Stops After
Sample, Peak Detect Single acquisition, one or two
channels simultaneously
Average, Envelope N acquisitions, one or two chan-
nels simultaneously, N is settable
from 2 to 256 or
Inputs
Input Coupling DC, AC, or GND
Input Impedance,
DC Coupled
1 MW ±1% in parallel with 25 pF ±2 pF
Maxi
m
u
m
Voltag
e
eeeS d
Overvoltage Category Maximum Voltage
B
e
tw
ee
n
S
ignal an
d
Common at Input
NC
CAT II Environment (refer to
page 1–13)
300 V
RMS
B
NC
CAT III Environment (refer to
page 1–13)
150 V
RMS
For steady-state sinusoidal waveforms, derate at 20 dB/decade above
100 kHz to 13 V
pk
at 3 MHz and above. Also, refer to Overvoltage
Category description on page 1–13.