Tektronix THS720P Work Light User Manual


 
Performance Verification
THS710A, THS720A, THS730A & THS720P Service Manual
4–17
3. Set the leveled sine wave generator frequency to 200 MHz if you are
checking a THS730A or to 100 MHz if you are checking a THS710A,
THS720A or THS720P.
4. Set the TekScope instrument VOLTS/DIV to 500 mV/div.
5. Set the leveled sine wave generator output level to approximately 50 mV
p-p
so that the measured amplitude is approximately 500 mV. (The measured
amplitude can fluctuate around 500 mV.)
6. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
7. Change the TekScope instrument setup using the following steps:
SCOPE TRIGGER Slope \ (falling edge)
8. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.