Tektronix THS720P Work Light User Manual


 
Performance Verification
THS710A, THS720A, THS730A & THS720P Service Manual
4–19
4. For each range, set the DC voltage source output to the level listed below,
and then compare the meter reading to the accuracy limits.
Range
DC Voltage Source
Output Level
Accuracy Limits
400 mV 60 mV 59.2 mV to 60.8 mV
400 mV 360 mV 357.7 mV to 362.3 mV
4 V 3.6 V 3.577 V to 3.623 V
40 V 36 V 35.77 V to 36.23 V
400 V 360 V 357.7 V to 362.3 V
880 V 792 V 783 V to 801 V
5. Set the DC voltage source output level to 0 V.