Agilent Technologies E9300 Stud Sensor User Manual


 
Getting Started
17
First check the section labelled
'635HYLVLRQ. Release A.01.11 or later is
required. If your power meter has an earlier release, please contact your
nearest Service Office (listed on page -70) to arrange an upgrade.
Next check the section labelled
0DLQ):5HY. Release A1.04.00 or later is
required for single channel meters; release A2.04.00 or later is required for dual
channel meters. For E9300 power sensors with suffix ‘B’ or ‘H’, firmware
revision A1.06.00 or later is required for single channel meters; revision
A2.06.00 or later is required for dual channel meters. If your power meter has
an earlier release, please contact your nearest Agilent Service Office (listed on
page -70) to arrange an upgrade.
Note You can carry out the firmware upgrade yourself if your power meter
has the required. Access http://www.agilent.com/find/powermeters
and click onthe link:
EPM Series E4418B Single-Channel Power Meter” or
EPM Series E4419B Dual-Channel Power Meter”. Click the
Software, Firmware and Drivers” link and follow the downloading
instructions.
Interconnections and Calibration
Connect one end of an Agilent 11730 series sensor cable to the Agilent E-series
E9300 power sensor and connect the other end of the cable to the power
meter’s channel input. Allow a few seconds for the power meter to download
the power sensor’s calibration table.
Caution The Agilent 9304A Sensor is DC coupled. DC voltages in excess of the
maximum value (5 Vdc) can damage the sensing diode.
Note Ensure power sensors and cables are attached and removed in an
indoor environment.
To carry out a zero and calibration cycle as requested by the power meter
proceed as follows:
Ensure the Agilent E-series E9300 power sensor is disconnected from
any signal source.
When calibrating Agilent E-series E9300B or E9301B sensors, first
remove the attenuator.