Agilent Technologies E9300 Stud Sensor User Manual


 
E9300/1B and H Power Sensor Specifications
48
Figure 17 E9300B Typical Power Linearity at 25°C, after zero and
calibration with associated Measurement Uncertainty
See Note on page -49.
E9300/1B -6 to 0 dBm 0 to 10 dBm 10 to 20 dBm 20 to 26 dBm
Measurement
Uncertainty
±0.65% ±0.55% ±0.45% ±0.31%
-10 -5 0 5 10 15 20 25 30
1
0
-1
0.8
0.6
0.4
0.2
-0.2
-0.4
-0.6
-0.8
Power Level (dBm)
% Error