Table 5: Power Sensor SWR and Reflection Coefficient for the Agilent 9304A
Table 21 Power Sensor SWR and Reflection Coefficient for the Agilent E9300B
Table 22 Power Sensor SWR and Reflection Coefficient for the Agilent E9301B
Table 23 Power Sensor SWR and Reflection Coefficient for the Agilent E9300H
Table 24 Power Sensor SWR and Reflection Coefficient for the Agilent E9301H
Frequency System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
9 kHz to 2 GHz ±0.010 0.061
2 GHz to 6 GHz ±0.010 0.087
Frequency System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
10 MHz to 8 GHz ±0.010 0.057
8 GHz to 12.4GHz ±0.010 0.078
12.4 GHz to 18 GHz ±0.010 0.107
Frequency System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
10 MHz to 6 GHz ±0.010 0.057
Frequency System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
10 MHz to 8 GHz ±0.010 0.070
8 GHz to 12.4GHz ±0.010 0.111
12.4 GHz to 18 GHz ±0.010 0.123
Frequency System Rho
Uncertainty
Actual
Measurement
Maximum
Rho
10 MHz to 6 GHz ±0.010 0.070