E9300/1/4/A Power Sensor Specifications
36
Figure 11 Typical Power Error Introduced in an Agilent E9304A
power sensor by DC Voltage
Power Linearity
After Zero and Calibration at ambient environmental conditions.
Power Level
Linearity
25°C ±10°C
Linearity
0°C to 55°C
-60 dBm to -10 dBm ±3.0% ±3.5%
-10 dBm to 0 dBm ±2.5% ±3.0%
0 dBm to +20 dBm ±2.0% ±2.5%