Power Meter Configuration Changes
The Agilent EPM series power meter recognizes when an Agilent E-series
E9300 power sensor is connected. The sensor calibration data is automatically
read by the power meter. In addition, the HP E-series E9300 power sensors
change the auto-averaging settings used by the power meter. These are also
automatically configured.
Figure 4 Auto-averaging Settings
Note These values are valid only for the power meter channel connected to
the Agilent E-series E9300 power sensor and only while the sensor is
connected. Averaging settings can also be manually configured. Refer
to “Achieving Stable Results with TDMA Signals” on page 25 if
required.
-4 dBm
-10 dBm
Sensor Power
Minimum
Sensor Dynamic Range
10 dBm
2 dBm
1234
1114
Resolution Setting
Number of Averages
11416
11
8
32
1 4 16 128
11664128
Maximum
-40 dBm
-50 dBm
-20 dBm
-30 dBm
1114
11216
121664
4 16 128 256
32 64 256 256
Upper Range
Lower Range
Sensor Power
6 dBm
0 dBm
20 dBm
12 dBm
-30 dBm
-40 dBm
-10 dBm
-20 dBm
26 dBm
20 dBm
40 dBm
32 dBm
-10 dBm
-20 dBm
10 dBm
0 dBm
E9300/1B
E9300/1HE9300/1/4A