Agilent Technologies E9300 Stud Sensor User Manual


 
General Information
60
General Information
This chapter contains information about general maintenance, performance
tests, troubleshooting and repair of Agilent E-series E9300 power sensors.
Cleaning
Use a clean, damp cloth to clean the body of the Agilent E-series E9300 power
sensor.
Connector Cleaning
Caution The RF connector beads deteriorate when contacted by hydrocarbon
compounds such as acetone, trichloroethylene, carbon tetrachloride,
and benzene.
Caution Clean the connector only at a static free workstation. Electrostatic
discharge to the center pin of the connector will render the power
sensor inoperative.
Keeping in mind its flammable nature; a solution of pure isopropyl or ethyl
alcohol can be used to clean the connector.
Clean the connector face using a cotton swab dipped in isopropyl alcohol. If
the swab is too big use a round wooden toothpick wrapped in a lint free cotton
cloth dipped in isopropyl alcohol. Refer to Agilent Application Note 326,
Principles of Microwave Connector Care (5954-1566) or Microwave Connector
Care (08510-90064) for proper cleaning methods.