Agilent Technologies E9300 Stud Sensor User Manual


 
E9300/1/4/A Power Sensor Specifications
35
Figure 10 Typical SWR 9 kHz to 6 GHz (25°C ±10°C) E9304A
Maximum Power
+25 dBm (320 mW) average
+33 dBm peak (2 W) <10µs
Maximum DC Voltage
The Agilent E9304A sensor is dc coupled. DC coupling of the input allows
excellent low frequency coverage. However, the presence of dc voltages mixed
with the signal will have an effect on the accuracy of the power measurement
(see graph below).
Caution DC voltages in excess of the maximum value (5 V) can damage the
sensing diode.
Maximum dc voltage: 5 Vdc (E9304A only)