E9300/1B and H Power Sensor Specifications
52
Error
Offset at Switch Point ≤±0.5% (≤±0.02 dB) typical
Switching Point Hysteresis 0.5 dB typical
E9300/1B
Conditions
(RH)
1
1. RH is the abbreviation for Relative Humidity.
Zero Set
Zero Drift
2
2. Within 1 hour after zero set, at a constant temperature, after a 24 hour
warm-up of the power meter with sensor connected.
Measurement
Noise
3
3. The number of averages at 16 for Normal mode and 32 for x2 mode, at a constant
temperature, measured over a one minute interval and two standard deviations.
Lower Range 15% to 75% 500 nW 150 nW 700 nW
(-30 to
+20 dBm)
75% to 95% 500 nW 4 µW 700 nW
Upper Range 15% to 75% 500 µW150 µW 500 µW
(+20 to
+44 dBm)
75% to 95% 500 µW 3 mW 500 µW
E9300/1H
Lower Range 15% to 75% 5 nW 1.5 nW 7 nW
(-50 to 0 dBm)
75% to 95% 5 nW 40 nW 7 nW
Upper Range 15% to 75% 5 µW 1.5 µW5 µW
(0 to +30 dBm) 75% to 95% 5 µW 30 µW5 µW