8 - 11
ENTRY OF SYSTEM PARAMETERS
ETC00265(4) Series 100 e 02/2004
END OF RANGE VALUE
8.13 End of Range Value
The operator is here offered the opportunity to introduce a full scale range for the analog signal
output.
Example:
For an analyzer concentration range of 0 - 25 % it is desired to measure only concentrations in
the range 0 - 15 %. If the operator enters here the value 15 %, the analog signal outputs of
10 (1) V / 20 mA will then correspond to a gas concentration of 15 %.
The displayed values are not affected.
Use the keys
and for the entry.
Entry possibility for channel 2
Use the keys
and for the entry.
Note:
The specifications of the analyzer written in the data sheet are only for OFS. = 0 and
END = full - scale range set in our factory !
It is part of customer to enter logical values for OFS. and END !